Whirl-Pack Write-On Whirl-Pak Collection Bags - Box of 100 - C32520N
- ... 1/8” ID tubing for filling and emptying the bag. The septum can be used for removing the sample using a syringe. Fittings are available in polypropylene, stainless steel, and Teflon..
- Guidelines: 1. Store Tedlar bags in a clean area away from potential contamination sources. 2. Tedlar bags are designed for use at atmospheric pressure. Over pressurization (003e2 psi) can cause the bags to burst. 3. Always ensure the bag, fitting, and tubing materials are appropriate for the compounds being sampled. 4. CEL Scientific does not recommend reusing Tedlar bags in most cases..
- To fill the bag: 1. Connect the appropriate tubing (1/8” ID) from a sampling source to the stem on the side of the fitting. 2. Open the valve by holding the stem and turning the top portion of the fitting counterclockwise one turn. 3. When the bag is full (approx. 90% capacity), close the valve by holding the stem and turning the top portion of the fitting clockwise until it is finger tight..
- To withdraw the sample using a syringe: 1. Insert the needle into the red septum on the top of the fitting. 2. Withdraw the sample and remove the needle, the septum will reseal itself. (Do not allow the needle to pierce the Tedlar film on the back side of the bag).